پمپ نمونه بردار فردی SKC مدل PCXR8 (( قابل برنامه ریزی ))
توضیحات محصول
224-PCXR8 Air Sampling Pump Specifications
Operating Range | 1000 to 5000 ml/min (UL Listed model) |
Weight | 34 oz (964 gm) |
Dimensions | 5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.8 cm) |
Compensation Range | 1000 to 2500 ml/min at 40 inches water back pressure 3000 ml/min at 35 inches water back pressure 4000 ml/min at 20 inches water back pressure 5000 ml/min at 10 inches water back pressure |
Flow Control | Holds constant flow to ± 5% of set-point |
Run Time: Battery | 8 Hours minimum at 4000 ml/min and 20 inches water back pressure;dependent on media used. |
Run Time: Mains Adapter | 9999 minutes (6.8 days). Pump will shut off as run time cannot exceed timer display (see Timer Display Range below) |
Timer Display Range: | 1 to 9999 minutes (6.8 days). Pump will shut off at 9999 minutes. To reset, restart the pump. |
Flow Indicator | Built-in rotameter with 250-ml division; scale marked at 1, 2, 3, 4, and 5 L/min |
Charging Time (varies with capacity and level of charge) | 6 to 8.5 hrs with PowerFlex charger |
Power Supply | 6.0 V plug-in NiCad battery pack, rechargeable, 2.0-Ah capacity |
Intrinsic Safety | UL Listed for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and Class III, Temperature Code T3C |
Operating Temperature | 32 to 113 F (0 to 45 C) |
Storage Temperature | -40 to 113 F (-40 to 45 C) |
Charging Temperature | 50 to 113 F (10 to 45 C) |
Operating Humidity | 0 to 95% Relative |
Flow and Low Battery Fault | Fault shutdown on excessive back pressure or low battery with LCD indicator and time display retention |
Battery Test | LCD indicator verifies battery condition prior to sampling |
Time Display | LCD indicator displays sampler run time in minutes |
Delay On | Allows user to select minutes to delay start of test up to 9999 minutes (6.8 days) |
Timed Shutdown | Allows user to select minutes of operation before automatic shutdown. Timed shutdown maximum is 9999 minutes (6.8 days) |
Timing Accuracy | ± 0.05% (± 45 sec/day) |
Intermittent Sampling | Programmable to allow user to extend short-term samples over an extended period of time to meet time weighted average (TWA) requirements with a reduced number of samples. Elapsed time maximum is 9999 minutes (6.8 days), at which time the sample pump shuts down. |
Sampling Pause | Allows user to temporarily halt sampling without loss of timing data. Restart does not require resetting time. |
Multiple-tube Sampling | Built-in constant pressure regulator allows user to take up to four simultaneous samples with rates up to 500 ml/min each (total combined flow 1350 ml/min maximum) using the optional adjustable low flow holder accessory |
RFI/EMI Shielding | Complies with requirements of EN 55022, FCC Part 15 Class B, EN 50082-1 Frequency range of the radiated susceptibility test was 27 to 1000 MHz CE marked |
Additional Specifications | Built-in fluid/particulate trap External see-through filter housing Tamper-resistant control cover Tamper-resistant compensation adjustments Built-in accessory mount Stainless steel belt clip External exhaust port All non-plastic parts made of corrosion-resistant materials RFI and EMI-shielded case |