پمپ نمونه بردار فردی SKC مدل PCXR8 (( قابل برنامه ریزی ))

پمپ نمونه بردار فردی SKC مدل PCXR8 PCXR8 Air Sampling Pump (( قابل برنامه ریزی ))

توضیحات محصول

224-PCXR8 Air Sampling Pump Specifications

Operating Range

1000 to 5000 ml/min (UL Listed model)
1000 to 3000 ml/min (MSHA-approved model)
(5 to 500 ml/min requires the 224-26 Series Adjustable Low Flow Holder)

Weight34 oz (964 gm)
Dimensions5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.8 cm)
Compensation Range1000 to 2500 ml/min at 40 inches water back pressure
3000 ml/min at 35 inches water back pressure
4000 ml/min at 20 inches water back pressure
5000 ml/min at 10 inches water back pressure
Flow ControlHolds constant flow to ± 5% of set-point
Run Time: Battery8 Hours minimum at 4000 ml/min and 20 inches water back pressure;dependent on media used.
Run Time: Mains Adapter9999 minutes (6.8 days). Pump will shut off as run time cannot exceed timer display (see Timer Display Range below)
Timer Display Range:1 to 9999 minutes (6.8 days). Pump will shut off at 9999 minutes. To reset, restart the pump.
Flow IndicatorBuilt-in rotameter with 250-ml division; scale marked at 1, 2, 3, 4, and 5 L/min
Charging Time
(varies with capacity and level of charge)
6 to 8.5 hrs with PowerFlex charger
Power Supply6.0 V plug-in NiCad battery pack, rechargeable, 2.0-Ah capacity
Intrinsic Safety

UL Listed for Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2, Groups E, F, G; and Class III, Temperature Code T3C
ATEX-approved model is available;

Operating Temperature32 to 113 F (0 to 45 C)
Storage Temperature-40 to 113 F (-40 to 45 C)
Charging Temperature50 to 113 F (10 to 45 C)
Operating Humidity0 to 95% Relative
Flow and Low Battery FaultFault shutdown on excessive back pressure or low battery with LCD indicator and time display retention
Battery TestLCD indicator verifies battery condition prior to sampling
Time DisplayLCD indicator displays sampler run time in minutes
Delay OnAllows user to select minutes to delay start of test up to 9999 minutes (6.8 days)
Timed ShutdownAllows user to select minutes of operation before automatic shutdown. Timed shutdown maximum is 9999 minutes (6.8 days)
Timing Accuracy± 0.05% (± 45 sec/day)
Intermittent SamplingProgrammable to allow user to extend short-term samples over an extended period of time to meet time weighted average (TWA) requirements with a reduced number of samples. Elapsed time maximum is 9999 minutes (6.8 days), at which time the sample pump shuts down.
Sampling PauseAllows user to temporarily halt sampling without loss of timing data. Restart does not require resetting time.
Multiple-tube SamplingBuilt-in constant pressure regulator allows user to take up to four simultaneous samples with rates up to 500 ml/min each (total combined flow 1350 ml/min maximum) using the optional adjustable low flow holder accessory
RFI/EMI ShieldingComplies with requirements of EN 55022, FCC Part 15 Class B, EN 50082-1
Frequency range of the radiated susceptibility test was 27 to 1000 MHz
CE marked
Additional SpecificationsBuilt-in fluid/particulate trap
External see-through filter housing
Tamper-resistant control cover
Tamper-resistant compensation adjustments
Built-in accessory mount
Stainless steel belt clip
External exhaust port
All non-plastic parts made of corrosion-resistant materials
RFI and EMI-shielded case