Product characteristics |
---|
Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 |
Measuring range [l/min] | 0...10 |
Application |
---|
Special feature | Gold-plated contacts |
Installation | T-pieces; ((DIN 2353): QL 18-18-18 (NW 15 mm) / QL 22-18-22 (NW 19 mm) / QL 28-18-28 (NW 24 mm)) |
Media | water |
Medium temperature [°C] | 0...80 |
Pressure rating [bar] | 30 |
Electrical data |
---|
Operating voltage [V] | 20...28 DC |
Current consumption [mA] | < 80 |
Protection class | III |
Reverse polarity protection | yes |
Power-on delay time [s] | 10 |
Inputs / outputs |
---|
Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 |
Outputs |
---|
Total number of outputs | 2 |
Output signal | switching signal; analogue signal; (configurable) |
Electrical design | PNP/NPN |
Number of digital outputs | 2 |
Output function | normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V] | 2 |
Permanent current rating of switching output DC [mA] | 250; (per output) |
Number of analogue outputs | 1 |
Analogue current output [mA] | 4...20; (scalable) |
Max. load [Ω] | 500 |
Analogue voltage output [V] | 0...10; (scalable) |
Min. load resistance [Ω] | 2000 |
Short-circuit protection | yes |
Type of short-circuit protection | pulsed |
Overload protection | yes |
Measuring/setting range |
---|
Measuring range [l/min] | 0...10 |
Display range [l/min] | 0...12 |
Set point SP [l/min] | 0.1...12 |
Reset point rP [l/min] | 0...11.9 |
Analogue start point ASP [l/min] | 0...8 |
Analogue end point AEP [l/min] | 4...12 |
In steps of [l/min] | 0.1 |
Accuracy / deviations |
---|
Flow monitoring |
---|
Repeatability [% of the measured value] | 3 (Q < 30%) / 7 (Q < 100%) |
Temperature gradient [K/min] | 200 |
Accuracy (in the measuring range) | ± 10 |
Response times |
---|
Flow monitoring |
---|
Response time [s] | 5; (10%...90%) |
Delay time programmable dS, dr [s] | 0...50 |
Software / programming |
---|
Parameter setting options | hysteresis / window; normally open / normally closed; switching logic; current/voltage output; inside pipe diameter; start-up delay; adaptation of displayed values; display can be rotated and switched off; Display unit |
Operating conditions |
---|
Ambient temperature [°C] | -20...60 |
Protection | IP 67 |
Tests / approvals |
---|
EMC | EN 61000-4-2 ESD | 4 kV CD / 8 kV AD | EN 61000-4-3 HF radiated | 10 V/m | EN 61000-4-4 Burst | 2 kV | EN 61000-4-6 HF conducted | 10 V |
|
MTTF [years] | 184 |
Mechanical data |
---|
Weight [g] | 330.5 |
Materials | stainless steel (1.4404 / 316L); PBT; PC; PA; EPDM/X; FKM |
Materials (wetted parts) | stainless steel (1.4404 / 316L); O-ring: FKM 70 Shore A |
Displays / operating elements |
---|
Display | switching status | 2 x LED, red | measured values | 7-segment LED display | programming | 7-segment LED display |
|
Display unit | l/min; gpm; l/h; gph; m³/h |
Remarks |
---|
Remarks | |
Pack quantity | 1 pcs. |
Electrical connection |
---|
Connection | Connector: 1 x M12; Contacts: gold-plated |