Standard |
718 Series |
|
Measure Pressure |
(internal sensor) |
Accuracy |
0.05% full scale |
Functions |
zero, min/max, hold, damp |
Note: |
Dry air, non-corrosives only |
|
|
718 1G |
|
Range: |
-1 PSI to +1 PSI, -68.9 mbar to 68.9 mbar (-6.89 kPa to 6.89 kPa) |
Resolution: |
0.0001 psi, 0.001 mbar, 0.0001 kPa |
Over Pressure: |
5x Full Scale |
Functions: |
Zero, Min, Max, Hold, Damp |
|
|
718 30G |
|
Range: |
-12 PSI to 30 PSI, (-850 mbar to 2 bar, -85 to 206.84 kPa) |
Resolution: |
0.001 psi, 0.1 mbar, 0.01 kPa |
Over Pressure: |
5x Full Scale |
Functions: |
Zero, Min, Max, Hold, Damp |
|
|
718 100G |
|
Range: |
-12 PSI to 100 PSI, (-850 mbar to 7 bar, -85 to 689.48 kPa) |
Resolution: |
0.01 psi, 0.1 mbar, 0.01 kPa |
Over Pressure: |
2x Full Scale |
Functions: |
Zero, Min, Max, Hold, Damp |
|
|
718 300G |
|
Range: |
-12 PSI to 300 PSI, (-850 mbar to 20 bar, -85 to 2068.42 kPa ) |
Resolution: |
0.01 psi, 1 mbar, 0.1 kPa |
Over Pressure: |
375 PSI 25 bar |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Fluke 718 series via using 700 Series Pressure Modules |
Range: |
29 pressure modules, 0 – 1 in H20 to 10,000 psi, 2.5 mbar to 700 bar. Over pressure per pressure module specs |
Resolution: |
Per pressure module specs |
Accuracy: |
To 0.025 % of full span, per pressure module specs |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: media compatibility per pressure module specs |
|
Fluke 718 series via built in pump |
Range: |
-12 PSI or -850 mbar to full scale Supported Pressure Units; psi, in H20 (4 °C), in H20 (20 °C), cm H20 (4 °C), cm H20 (20 °C), bar, mbar, kPa, inHg, mmHg, kg/cm |
Loop Power: |
Range: 24 V dc |
|
Accuracy: +/- 10 % |
|
Note: Drive; 20 mA into 1000 Ω for battery > 6.8V; 700 Ω for battery 5.8 to 6.8 V |
mA Measure: |
Range: 0 mA to 24 mA |
|
Resolution: 0.001 mA |
|
Accuracy: 0.015 % + 1 count |
|